Pressure Controlled & Airtight Split Coin Cell
This is a solid-state split-able coin cell set with a PEEK (Optional Silicon Nitride, Si3N4, for heating) sleeve with optional inner diameters (ID) of 10mm, 20mm, and 30mm and outer diameters of 40mm continued temperature up to 300°C. The split cell can be placed in a lab press with pressure up to 150 MPa. It is designed to research the effect of pressure on solid-state battery performance, and it can also be used for easy coin cell assembling without any screw and crimping tools.
Specifications
Device composition:
Transfer fixture, insulation mold with Si3N4 sleeve including heating jacket, a temperature controller, pressure sensor, and digital display meter.
Size:
Die inner diameter: 10mm, 20mm, 30mm
Mold outer diameter: 40 mm
Total height: 85 mm
Temperature Range: up to 300°C
Pressure range: 0 – 5T
Sensitivity 1.0 – 2.0 (± 0.1 mv/v )
Voltage: 220V, AC Single phase
Structure and materials
1. The insulation mold is composed of a 10mm internal diameter pressing rod, a pressing piece, a gasket, a mold bottom, and the sleeve is made of silicon nitride.
2. Pushing rods (top and bottom) are made of stainless steel
3. The sleeve’s inner diameter is available with Ø10 mm, Ø20mm and Ø30mm
2. Easy coin cell assembling without any screw and crimping tools.
3. High working temperatures and pressure applications are available
4. To test a solid-state cell at heating conditions, please select a heatable Si3N4 solid-state cell combination that includes a heating jacket and a temperature control unit.
Transfer device
1. 3 M8 screw rods, 90mm diameter upper cover, pressure plate and base.
2. The external diameter of the device is 90mm, and the total height of the screw rod is
146mm.
Pressure sensor
1. Range: 0-5T.
2. Sensitivity 1.0-2.0 ± 0.1mv/v
3. Working temperature range – 20 -+70 ℃
Digital display meter
1. Digital display KG, range 0-5000KG, (adjustable)
2. Voltage: 220V
Die pressure test
The pressure can reach 300-500 mpa (the inner diameter of the insulation mold
varies with the inner diameter). Press to 1 ton, the pressure is about 125 mpa
Repeat error (Newtonian liquid): ±1%
Net weight: 6.8 Kg
The climate chamber / constant temperature and humidity test chamber used for testing cells and packs at different temperature conditions, damp heat and to check reliability test at different environmental conditions.
The equipment prohibits:
Model: BS-CTH150L
Specifications:
Model: BS-CTH1000L
Specifications: