Description:
High-Resolution Automated Raman Spectrograph is a unique high-aperture double dispersive Monochromator-spectrograph whose compact design combines high spectral resolution essential for long-focus instruments and extremely low loss of light which is an essential requirement for Raman spectroscopy, as well as for any other low-light-level application.
Features:
Applications:
Specifications:
Spectral range | Typical 180 – 4800nmUp to 40µm when the appropriate gratings are used | |||||
F/Number | 1 : 5.5 | |||||
Effective focal length | 833 mm | |||||
Ports | Two inputs and two outputs | |||||
Flat field | 30 mm x 10 mm | |||||
Imaging | Option. Available for the lateral output port | |||||
Diffraction gratings | 70 mm x 70 mm x 10 mm, Automated turret with three pairs | |||||
Lines/mm | 2400 | 1800 | 1200 | 900 | 600 | 300 |
Blaze wavelength, nm | 225 | 400 | 600 | 600 | 1000 | 2000 |
Spectral range, nm | 190 – 600 | 253 – 800 | 380 -1200 | 506 – 1600 | 760 – 2400 | 1500 – 4800 |
Scanning range, nm | 0 – 640 | 180 – 850 | 0 – 1280 | 0 – 1710 | 0 – 2560 | 0 – 5120 |
Wavelength accuracy, nm | ±0.03 | ±0.04 | ±0.06 | ±0.08 | ±0.12 | ±0.24 |
Wavelength reproducibility,nm | ±0.006 | ±0.008 | ±0.012 | ±0.016 | ±0.024 | ±0.048 |
Step of grating rotation, nm | 0.0017 | 0.0023 | 0.0035 | 0.0046 | 0.007 | 0.014 |
Scanning speed, nm/s | 5,3 | 7 | 10.5 | 14 | 23 | 42 |
Entrance/exit slits | Automatic and manual adjustment | |||||
Filter wheel | Automatic switching | |||||
Max number of filters | 8 | |||||
Standard number of filters | 5 | |||||
Filter size | 20 mm | |||||
Light aperture | 18 mm |
Description:
The M266 Automated Monochromator-spectrograph is general-purpose high-aperture instrument with high line quality and low stray light performance, featuring 284 mm focus and 30 mm x 10 mm flat field.
Features:
Applications:
Specifications
Spectral range | – Typical 180 – 3600nm- Extended up to 40µm | ||||||||
F/Number | 1 : 3.8 | ||||||||
Focal length | 284 mm | ||||||||
Flat field | 30 mm x 10 mm | ||||||||
Imaging | Option. Available for the both output ports simultaneously | ||||||||
Gratings | 50 mm x 50 mm x 10 mm, One grating or a turret with 4 gratings | ||||||||
Lines/mm | 2400 | 1800 | 1200 | 600 | 400 | 300 | |||
Blazing wavelength, nm | 225 | 270 | 400 | 750 | 1000 | 800 | 1700 | 1500 | 2000 |
Spectral range, nm | 180-450 | 180-600 | 265-900 | 500-1500 | 660-1800 | 530-1600 | 1130-2700 | 1000-3000 | 1330-3600 |
Wavelength accuracy, nm | ±0.05 | ±0.067 | ±0.1 | ±0.2 | ±0.3 | ±0.4 | |||
Wavelength reproducibility,nm | ±0.013 | ±0.017 | ±0.025 | ±0.05 | ±0.075 | ±0.1 | |||
Step of grating rotation, nm | 0.007 | 0.009 | 0.013 | 0.026 | 0.04 | 0.05 | |||
Scanning speed, nm/s | 12.5 | 16.7 | 25 | 50 | 75 | 100 | |||
Entrance/exit slits | Automatic and manual adjustment | ||||||||
Filter wheel | Automatic switching | ||||||||
Max number of filters | 8 | ||||||||
Standard number of filters | 5 | ||||||||
Filter size | 20 mm | ||||||||
Light aperture | 18 mm |
Description:
The M150 Multi-Purpose Compact Monochromator-Spectrograph is a low-cost short-focus, flexible and multifunctional instrument, featuring a unique combination of performance specifications and equipped with all the options essential for large-size spectrographs.
Features:
Applications:
Specifications:
Spectral range | -Typical 190 – 3600nm-Extended up to 15µm | |||||||||
F/Number | 1 : 3.6 | |||||||||
Focal length | 142 mm | |||||||||
Flat field | 30 x 10 mm | |||||||||
Imaging | Option. Available on both output ports | |||||||||
Diffraction gratings | 40 mm x 40 mm x 10mm, Turret with 3 gratings | |||||||||
Lines/mm | 1800 | 1200 | 900 | 600 | 600 | 400 | 300 | 200 | 100 | 100 |
Blaze wavelength, nm | 270 | 400 | 600 | 750 | 1000 | 800 | 1500 | 1500 | 2000 | 500 |
Spectral range, nm | 190-540 | 265-800 | 400-1200 | 500-1500 | 660-1800 | 530-1600 | 1000-3000 | 1000-3000 | 1330-4000 | 330-1000 |
Entrance slits | Automatic and manual adjustment | |||||||||
Filter wheel | Automatic switching | |||||||||
Max number of filters | 8 | |||||||||
Standard number of filters | 5 | |||||||||
Filter size | 20 mm | |||||||||
Light aperture | 18 mm |
Description:
The MSA-130 Double Additive/Subtractive Monochromator is a compact double Monochromator with enhanced functionality and low stray light.
The key feature of the MSA-130 is its unique ability to operate in both dispersion additive and subtractive dispersion modes.
The MSA-130 operation modes are changed over by manual switching of a two-position selector switch from the position “Addition” to “Subtraction” and vice versa without any additional alignment by the user. The instrument is multifunctional, compact, easy-to-use and reliable.
Features:
High Throughput
Specifications:
Focal length | |||
The first Monochromator, collimating mirror | 130 mm | ||
focusing mirror | 142 mm | ||
The second Monochromator, collimating mirror | 142 mm | ||
Focusing mirror | 130 mm | ||
Ports | 1 input, 1 output | ||
F/Number | 1 : 4.5 | ||
Entrance & Exit slits, Micrometric | |||
Slit Width & Height | 0 to 2 mm (W) & 12 mm(H) | ||
Parallelism | +/- 1 μm | ||
Micrometer Reading Accuracy | +/- 1 μm | ||
Intermediate slit, Micrometric | |||
Slit width & Height | 0 to 5 mm (W) & 12 mm (H) | ||
Parallelism | +/1 μm | ||
Micrometer Reading Accuracy | +/1 μm | ||
GRATINGS | |||
Number of Lines per mm | 1200 | 600 | 300 |
Grating Size, mm | 25*25*8 | 25*25*8 | 25*25*8 |
Blazing Wavelength, nm | 280 | 600 | 1500 |
Usable Wavelength Range, nm | 190- 650 | 385 – 1300 | 770-2600 |
Mechanical Range, nm | 0 – 760 | 0 -1500 | 0 – 3000 |
WAVELENGTH | |||
Reciprocal Dispersion of the first Monochromator | 5.8 nm/mm | 11.6 nm/mm | 23.2 nm/mm |
Wavelength Accuracy | +/- 0.15 nm | +/- 0.25 nm | +/- 0.5 nm |
Wavelength Repeatability | +/- 0.05 nm | +/- 0.1 nm | +/- 0.2 nm |
Wavelength Step Size | 0.01 nm | 0.02 nm | 0.04 nm |
Wavelength Scan Speed | 19 nm/s | 38 nm/s | 76 nm/s |
Dispersion Addition Mode | |||
Reciprocal Dispersion | 3.0 nm/mm | 6.0 nm/mm | 12.0 nm/mm |
Spectral Resolution | 0.07 nm | 0.14 nm | 0.28 nm |
Width of the Selected Spectral Range | 0.2 – 6 nm | 0.4 – 12 nm | 0.8 – 24 nm |
Dispersion Subtraction Mode | |||
Reciprocal Dispersion | 0 nm/mm | 0 nm/mm | 0 nm/mm |
Width of the Selected Spectral Range | 0.2 – 40 nm | 0.4 – 80 nm | 0.8 – 160 nm |
Width of 25μm Entrance Slit Imagewith 2.5mm intermediate slit | 32 µm | 32 µm | 32 µm |
With 5.0mm intermediate slit | 43 µm | 43 µm | 43 µm |
Stray Light | 6 * 10– 9 |
Description:
The ML44 High-Aperture Compact Monochromator is a compact high-aperture short-focus Monochromator operating in the spectral range of 190-1200nm. Complemented by a light source, the ML44 can be successfully used as a monochromatic illuminator for fluorescence excitation.
Features:
Applications:
Specifications:
Spectral range, nm | 190-1200 | |
Focal length of the collimator and camera lenses, mm | 44.2 | |
F/Number | 1 : 2.9 | |
Gratings | ||
Size, mm | 25 * 20 * 6 | |
Lines/mm | 1200 | 600 |
Blazing wavelength, nm | 270 | 700 |
Reciprocal linear dispersion, nm/mm | 18.7 | 34.7 |
Operating spectral range, nm | 190-600 | 450-1200 |
Wavelength of automatic change of grating, nm | 600 (Program controlled) | |
Tuning step by angle, deg | 0,01875 | |
Tuning step by spectrum, nm | 0.125 | 0.25 |
Wavelength accuracy, nm | ±1.0 | ±1.5 |
Wavelength repeatability, nm | ±0.25 | ±0.5 |
Scanning speed, nm/s | 200 | 400 |
Time required for wavelength tuning, s | 0.001 (Min.), 38.4 (Max.) | |
Entrance/exit slits (changeable) | ||
Slit width, mm | 0.02, 0.05, 0.1, 0.25, 0.5, 1.0 | |
Slit height, mm | 3.0 |