Monochromator

High-Resolution Automated Raman Spectrograph
monochromator1

Description:

High-Resolution Automated Raman Spectrograph is a unique high-aperture double dispersive Monochromator-spectrograph whose compact design combines high spectral resolution essential for long-focus instruments and extremely low loss of light which is an essential requirement for Raman spectroscopy, as well as for any other low-light-level application.

Features:

      • Extremely low loss of light
      • The compact long-focus instrument with the unique high aperture
      • High-Resolution Automated Raman Spectrograph is in the Imaging mode
      • Three pairs of gratings
      • Two input and two output ports
      • Full computer control

Applications:

    • Raman Spectroscopy
    • Emission and Fluorescence Spectroscopy
    • Analytical Tasks Demanding High Resolution in the Range from UV to IR
    • Multi-Channel Spectroscopy

Specifications:

Spectral range Typical 180 – 4800nmUp to 40µm when the appropriate gratings are used
F/Number 1 : 5.5
Effective focal length 833 mm
Ports Two inputs and two outputs
Flat field 30 mm x 10 mm
Imaging Option. Available for the lateral output port
Diffraction gratings 70 mm x 70 mm x 10 mm, Automated turret with three pairs
Lines/mm 2400 1800 1200 900 600 300
Blaze wavelength, nm 225 400 600 600 1000 2000
Spectral range, nm 190 – 600 253 – 800 380 -1200 506 – 1600 760 – 2400 1500 – 4800
Scanning range, nm 0 – 640 180 – 850 0 – 1280 0 – 1710 0 – 2560 0 – 5120
Wavelength accuracy, nm ±0.03 ±0.04 ±0.06 ±0.08 ±0.12 ±0.24
Wavelength reproducibility,nm ±0.006 ±0.008 ±0.012 ±0.016 ±0.024 ±0.048
Step of grating rotation, nm 0.0017 0.0023 0.0035 0.0046 0.007 0.014
Scanning speed, nm/s 5,3 7 10.5 14 23 42
Entrance/exit slits Automatic and manual adjustment
Filter wheel Automatic switching
Max number of filters 8
Standard number of filters 5
Filter size 20 mm
Light aperture 18 mm

line

Automated Monochromator-Spectrograph
automatedmono

Description:

The M266 Automated Monochromator-spectrograph is general-purpose high-aperture instrument with high line quality and low stray light performance, featuring 284 mm focus and 30 mm x 10 mm flat field.

Features:

  • High aperture ratio, low stray light and perfect line quality
  • Two output ports
  • Completely automated
  • Automatic shutter

Applications:

  • Emission, Fluorescence and Raman Spectroscopy
  • Multi-channel Spectroscopy
  • Absorption, Reflection and Transmission Measurements
  • Analytical Tasks from UV to IR

Specifications

Spectral range – Typical 180 – 3600nm- Extended up to 40µm
F/Number 1 : 3.8
Focal length 284 mm
Flat field 30 mm x 10 mm
Imaging Option. Available for the both output ports simultaneously
Gratings 50 mm x 50 mm x 10 mm, One grating or a turret with 4 gratings
Lines/mm 2400 1800 1200 600 400 300
Blazing wavelength, nm 225 270 400 750 1000 800 1700 1500 2000
Spectral range, nm 180-450 180-600 265-900 500-1500 660-1800 530-1600 1130-2700 1000-3000 1330-3600
Wavelength accuracy, nm ±0.05 ±0.067 ±0.1 ±0.2 ±0.3 ±0.4
Wavelength reproducibility,nm ±0.013 ±0.017 ±0.025 ±0.05 ±0.075 ±0.1
Step of grating rotation, nm 0.007 0.009 0.013 0.026 0.04 0.05
Scanning speed, nm/s 12.5 16.7 25 50 75 100
Entrance/exit slits Automatic and manual adjustment
Filter wheel Automatic switching
Max number of filters 8
Standard number of filters 5
Filter size 20 mm
Light aperture 18 mm

line

Multi-Purpose Compact Monochromator-Spectrograph
multimono

Description:

The M150 Multi-Purpose Compact Monochromator-Spectrograph is a low-cost short-focus, flexible and multifunctional instrument, featuring a unique combination of performance specifications and equipped with all the options essential for large-size spectrographs.

Features:

  • Extremely high aperture ratio
  • Low stray light
  • Three diffraction gratings
  • High degree of automation

Applications:

  • Fluorescence and Raman Spectroscopy
  • As a Spectral Selection Device for use with a Streak-Camera
  • Multichannel Spectroscopy
  • Measurement of Absorption, Reflection and Transmittance from UV to IR range

Specifications:

Spectral range -Typical 190 – 3600nm-Extended up to 15µm
F/Number 1 : 3.6
Focal length 142 mm
Flat field 30 x 10 mm
Imaging Option. Available on both output ports
Diffraction gratings 40 mm x 40 mm x 10mm, Turret with 3 gratings
Lines/mm 1800 1200 900 600 600 400 300 200 100 100
Blaze wavelength, nm 270 400 600 750 1000 800 1500 1500 2000 500
Spectral range, nm 190-540 265-800 400-1200 500-1500 660-1800 530-1600 1000-3000 1000-3000 1330-4000 330-1000
Entrance slits Automatic and manual adjustment
Filter wheel Automatic switching
Max number of filters 8
Standard number of filters 5
Filter size 20 mm
Light aperture 18 mm

line

Double Additive/Subtractive Monochromator
doubleadditive

Description:

The MSA-130 Double Additive/Subtractive Monochromator is a compact double Monochromator with enhanced functionality and low stray light.

The key feature of the MSA-130 is its unique ability to operate in both dispersion additive and subtractive dispersion modes.

The MSA-130 operation modes are changed over by manual switching of a two-position selector switch from the position “Addition” to “Subtraction” and vice versa without any additional alignment by the user. The instrument is multifunctional, compact, easy-to-use and reliable.

Features:

  • Double Monochromator for Excellent Stray Light Rejection
  • Multifunctional: additive/dual dispersion & non-additive/subtractive dispersion
  • Compact & Cost-effective

High Throughput

Specifications:

Focal length
The first Monochromator, collimating mirror 130 mm
focusing mirror 142 mm
The second Monochromator, collimating mirror 142 mm
Focusing mirror 130 mm
Ports 1 input, 1 output
F/Number 1 : 4.5
Entrance & Exit slits, Micrometric
Slit Width & Height 0 to 2 mm (W) & 12 mm(H)
Parallelism +/- 1 μm
Micrometer Reading Accuracy +/- 1 μm
Intermediate slit, Micrometric
Slit width & Height 0 to 5 mm (W) & 12 mm (H)
Parallelism +/1 μm
Micrometer Reading Accuracy +/1 μm
GRATINGS
Number of Lines per mm 1200 600 300
Grating Size, mm 25*25*8 25*25*8 25*25*8
Blazing Wavelength, nm 280 600 1500
Usable Wavelength Range, nm 190- 650 385 – 1300 770-2600
Mechanical Range, nm 0 – 760 0 -1500 0 – 3000
WAVELENGTH
Reciprocal Dispersion of the first Monochromator 5.8 nm/mm 11.6 nm/mm 23.2 nm/mm
Wavelength Accuracy +/- 0.15 nm +/- 0.25 nm +/- 0.5 nm
Wavelength Repeatability +/- 0.05 nm +/- 0.1 nm +/- 0.2 nm
Wavelength Step Size 0.01 nm 0.02 nm 0.04 nm
Wavelength Scan Speed 19 nm/s 38 nm/s 76 nm/s
Dispersion Addition Mode
Reciprocal Dispersion 3.0 nm/mm 6.0 nm/mm 12.0 nm/mm
Spectral Resolution 0.07 nm 0.14 nm 0.28 nm
Width of the Selected Spectral Range 0.2 – 6 nm 0.4 – 12 nm 0.8 – 24 nm
Dispersion Subtraction Mode
Reciprocal Dispersion 0 nm/mm 0 nm/mm 0 nm/mm
Width of the Selected Spectral Range 0.2 – 40 nm 0.4 – 80 nm 0.8 – 160 nm
Width of 25μm Entrance Slit Imagewith 2.5mm intermediate slit 32 µm 32 µm 32 µm
With 5.0mm intermediate slit 43 µm 43 µm 43 µm
Stray Light 6 * 10– 9

 

 

line

High-Aperture Compact Monochromato Model-ML44
highmono

Description:

The ML44 High-Aperture Compact Monochromator is a compact high-aperture short-focus Monochromator operating in the spectral range of 190-1200nm. Complemented by a light source, the ML44 can be successfully used as a monochromatic illuminator for fluorescence excitation.

Features:

  • Extremely High Aperture
  • Absence of astigmatism at the output
  • Change slits on the fly
  • Compact size and low cost

Applications:

  • For investigation of detectors and light sources
  • For broadband fluorescence excitation sources
  • For spectroscopy education at colleges and universities
  • As part of other research and analytical equipment operating in the UV, visible and NIR spectral range

Specifications:

Spectral range, nm 190-1200
Focal length of the collimator and camera lenses, mm 44.2
F/Number 1 : 2.9
Gratings
Size, mm 25 * 20 * 6
Lines/mm 1200 600
Blazing wavelength, nm 270 700
Reciprocal linear dispersion, nm/mm 18.7 34.7
Operating spectral range, nm 190-600 450-1200
Wavelength of automatic change of grating, nm 600 (Program controlled)
Tuning step by angle, deg 0,01875
Tuning step by spectrum, nm 0.125 0.25
Wavelength accuracy, nm ±1.0 ±1.5
Wavelength repeatability, nm ±0.25 ±0.5
Scanning speed, nm/s 200 400
Time required for wavelength tuning, s 0.001 (Min.), 38.4 (Max.)
Entrance/exit slits (changeable)
Slit width, mm 0.02, 0.05, 0.1, 0.25, 0.5, 1.0
Slit height, mm 3.0